Accumulative Method Ion Sensor (AMIS)

Principle

The drain current flowing through semi-conductor will be determined by electronic density inside of the semi-conductor. The chemical ion change caused by biological reaction such as enzyme reaction will be transferred into semiconductor as electric field change. The electric field change causes polarization in semi-conductor, which change the current flowing through the gate channels.

Signal Accumulation

AMIS sensor stores the signal measured per measurement inside the sensor for amplification to improve S/N ratio, which realized high sensitive and precise measurement, while conventional ISFET sensor cannot achieve because of poor S/N ratio due to small signal.

Example

Various enzymatic reaction measurement example

Possible to detect various kinds of enzymatic reactions in sensitive

Activity measurement of Lipase

And detecting the resulting oleic acid hydrolysis of Olive oil due to the Lipoprotein Lipase

Measurements of Flavonoids in the beverage

Measurement of Creatinine

Detection example of Creatinine hydrolysis reaction due to Creatinine Deiminase

Conventional three types of enzymes are required  Easily it can be measured in a single enzyme

Detection of Pesticide Residue

 

The features of Micro Bioactivity Analyzer AMIS-101are shown in

Relevant Literature (English)

  • Simple and convenient measurement of enzyme reaction by high sensitive signal accumulation ISFET biosensor (AMIS Sensor) and micro bioactivity analyzer

    Biocatalysis and Agricultural Biotechnology (2012)